Dinamap v100 manual




















Page Parameter Level Functional Testing Take an oral or axillary temperature reading on yourself i. Verify that a predictive temperature begins once the probe is removed from its holster.

Place the probe in the holster after completion of the Temp cycle. Page Exergen Connect the cables prior to attaching to the monitor. Verify that an SpO reading is displayed within moments of attaching the sensor to either a simulator or to your finger.

Page Setup Allow the system to stabilize for 5 seconds it is normal to see some decrease in pressure at this point. Start the stopwatch and record the pressure value.

After 60 seconds record the pressure value. Page Pressure Transducer Verification Record and verify the pressure reading that appears in the Systolic window. Record and verify the pressure reading that appears in the Diastolic window. Turn unit off. Remove the air restriction. Press the Silence button. Temperature perform if equipped with Temp module At installation, GE recommends performing a temperature check.

Page Alaris Temperature Calibration Verification for instructions. Calibration is complete. If the monitor does not pass the calibration verification, contact GE Technical Support.

If this is not the case, contact GE Technical Support. Place the probe on your finger. With the monitor powered on, turn it so that the side with the printer is facing you. While grasping the side of the monitor, lift the printer door open by placing your thumb in the indented area and pulling. Replace the power cord, as necessary, with a regulatory-approved cord for the country of use.

Page 84 Repeat the steps in this procedure using the appropriate Test Body for any additional devices requiring safety testing. If measured reading is greater than the acceptance criteria below, the device under test fails. Contact GE Technical Support. Page 86 Repeat the steps in this procedure using the appropriate Test Body for any additional patient connectors requiring safety testing. Acceptance criteria All readings must be less than or equal to 5 mA.

Page Overview A systematic approach to the diagnosis of problems as well as a general understanding of the architecture, both hardware and software, of the monitor are essential to ensure successful troubleshooting of a device. GE recommends formal service training before repairs are attempted. These troubleshooting procedures combined with training provide the service technician with skills necessary to service and repair a monitor in the event of a malfunction.

Page Alarm Code Interpretation Alarm code interpretation General system error codes are listed in this section. If any other system or similar code appears, record the error message and report the failure to GE Technical Support. Page 95 Line Frequency LF mode setting. Processing Error Restart scanner. If unit is defective, it will have to be returned to the manufacturer. Page Ordering Parts This section of the manual provides parts lists for the monitor.

Parts lists should be used in conjunction with the other chapters of this manual. GE makes every effort possible to provide the most up-to-date reference documentation for your equipment. Page - Cable Assy - 3. GE does not make available schematics, assembly drawings, or bills of material beyond what is provided in this manual.

Attempting repair on a PCB or factory- sealed component is not recommended. Select applicable label per monitor configuration. Kit includes only one. For non-printer versions, the print key will need to be trimmed before installation.

See parts drawing on page for further details. Page Monitor Fascia Replacement Procedure Pry open the front face plate at the upper corners and remove the face plate. Remove the existing fascia.

Remove the tab at the top of the new fascia that includes the part number. Page Turn the fascia right side up and align the fascia with the bottom notch on the monitor. Press the fascia onto the monitor. Tuck the top part of the fascia under the top notch on the monitor.

Page Monitor Disassembly Procedure To avoid damage when disassembling the monitor, observe the standard precautions and procedures for handling static-sensitive components.

Main battery Remove 4 screws securing the instruction cards. Remove the battery compartment door and adhesive pad. Page Rear Case Remove remaining two screws from the bottom of the monitor. Set monitor upright. Carefully remove the front faceplate.

Remove 2 screws behind black overlay. Set monitor on its face and open the printer door. Carefully remove the rear case. Page Printer Lift the 2 black tabs and remove printer cable. Remove printer assembly. Unplug air hose from valve and filter. You have a cart already in session. What action do you want to perform? Continue Checkout. I Agree. View Cart Continue Shopping. You already have promotional items in your cart ready to checkout.

Adding this "request for a quote" item will require a Quote and prevent checkout. Do you want to continue? Cancel Continue. You already have a "request for a quote" item in your cart. Adding this promotional item will still require a Quote and prevent Checkout. Sign In. This form will auto-populate account information for registered users.

How would you like to proceed? Proceed as Guest Login. Sorry, your details do not match our records. Re-enter your account number, zipcode to try again. Please enter appropriate Account ID. Please enter appropriate zipcode. Sorry, there is an internal server error. Please try again after some time. Activation Email Sent Successfully.

Login Password Activate. Please enter Email Address Next. Remember Me Forgot Password? Service Shop has been migrated to a new platform to provide you with a better customer experience. Website by Evoluted New Media. V List Subscribe to our Mailing List for dinamapp, webinar annoucements, and software releases.

Carescape V is fully customisable, with a range of blood pressure cuff, temperature probe and SpO2 options to choose from. Purchase Options Carescape V is available with a range of blood pressure, SpO2 and temperature probe options. Remember Me Forgot Password? And if the battery is discharged, it maintains the data, and you can plug it dinamqp and continue to use the monitor.

It also includes the same advanced parameters and algorithms as other higher acuity GE monitors, ensuring measurement consistency across all care areas. Alaris Turbo Temp provides fast oral or rectal temperatures at the bedside, as well as axillary temperatures in monitor mode. The Dinamap vital signs monitor is lightweight and easy to transport, making it convenient to take from one patient room to another.

This extended battery life allows you to flexibly move the monitor from one patient room to the next for the course of a shift. The silver ion anti-microbial head kills microbes such as E-Coli, Staph and Listeria. Add to Wish List. And 3 Alaris Tri-Site thermometer provides fast oral, rectal and axillary temperatures at the bedside. This form will auto-populate account information for registered users. Infection Prevention GE supports your infection prevention efforts.

The fuses are mounted within the monitor. The monitor can also be configured with an optional printer and a rolling stand can also be added. GE Dinamap Pro V2.

The GE Carescape V patient monitor is a reliable, easy-to-use vital sign monitor that helps you care dinamp many patients efficiently.